NanoSpectralyzer Technical Specifications
- Fluorescence excitation laser lambda .................... 532, 638, 671 and 785 nm
- Fluorescence Geometry ............................................. High numerical aperture epifluorescence
- Fluorescence spectral range .................................... 900-1600 nm
- Near-IR detector type .................................................. 512 element TE-cooled InGaAs array
- Raman* excitation laser lambda .............................. 532 or 671 nm (choose when ordering)
- Raman spectral range ............................................... 150 to 2900 cm-1 shift
- Raman spectral resolution ....................................... 4 cm-1
- Raman detector type .................................................. 3648 pixel TE-cooled Si CCD
- Absorption light source .............................................. Stabilized tungsten-halogen lamp
- Absorption spectral range ......................................... 400-1600 nm
- Absorption spectral resolution ................................. 4 nm (NIR), 1 nm (vis)
- Absorption ceiling ....................................................... 3 AU (NIR and vis)
- Visible detector type ................................................... 2048 pixel Si CCD
- Absorbance noise (rms), near-IR ........................... <2 x 10-4 at 0 AU for 10 s integration
- Absorbance noise (rms), visible ............................. <5 x 10-4 at 0 AU for 10 s integration
- Minimum sample volume ......................................... 120 uL
- Data acquisition time (typical) .................................. 2 minutes for full set of spectra
- Power consumption ................................................... 75 W (excluding computer)
- Main optical module dimensions ........................... 12.3" W x 18.3" D x 7.7" H (310 x 465 x 195 mm)
- System weight ............................................................. 48 lbs/22 kg (excluding computer)
*Raman available with NS2 only